Description
Application
All-in-one Gamma Spectrometer with digital signal processing is applicable for HPGe detector spectrometric pulses registration, amplification, digital filtration and gamma-ray spectra registration. All-in-one Gamma Spectrometer is designed as one integrated monoblock being identical to standard laboratory detection unit with inbuilt preamplifier, however it has inbuilt MCA, HV supply, etc.
Features
- All-in-one Gamma Spectrometer is fast and easy to connect the computer using Plug&Play method. Start operating the device in no more than 120 seconds
- Power supply for all parts of the spectrometer is provided by only one USB cable (USB-2 port). Maximum current less than 300 mA, voltage +5V
- All basic technical parameters of All-in-one Gamma Spectrometer are identical to any regular laboratory device produced by BSI
- Cryostat geometries and modifications are available
COMPLETE SET
- Charge sensitive preamplifier with resistive feedback and cooled FET
- Amplifier with intermediate differentiation
- Digital MCA
- HV supply +/-5000 V (two options) for positive and negative polarity
- Detector temperature sensor
- LED indicating detector operating temperature
- Energy registration range switch between 40 keV – 3.0 MeV and 3 keV – 300 keV
- USB cable for connection with computer
SPECIFICATION
Parameter | Value |
---|---|
ADC | 14bit, 10 MSps |
Integral nonlinearity | <0.04% |
Temperature stability | TK = 50 ppm |
Digital signal processing | |
Automated and manual adjustment of trigger threshold | |
Fine gain adjustment (in steps of <0.05%) | 0.5 – 6.5 |
Shaping time (Integration time / rise time of filter multiplied by 2) adjustable in the range (flat top adjustable 0 … 15μs) | 0.1… 25.5μs |
Channel splitting | 128, 256, 512, 1k, 2k, 4k, 8k, 16k |
Differential nonlinearity (for 8k channels and 6μs shaping time) | <1% |
Automated PZC adjustment, detector decay time constants from 40μs to 1ms can be compensated | |
Double differential trigger filter | |
Modes of operation – PHA (pulse height analysis), result is a histogram of pulse height. – MCS (multi channel scaling): registers number of counts in subsequent time intervals. – MCS mode allows to acquire an integral PHA spectrum in parallel to the MCS acquisition. – LLD / ULD settings are supported in both PHA and MCS mode. | |
Energy resolution for HPGe detector with efficiency of 7%* at 122 keV at 1.33 MeV |
0,820 keV 1,76 keV |
Throughput >100kcps in memory at 150kcps input rate and 0.2μs shaping time and 1.4μs flattop | |
High Voltage Supply (programmable) | +/- 4500V, 0.1mA max |
Power supply: USB, 4.5 V … 5.25 V 280mA at full operation and high voltage 4500 V | |
Computer Interface / Connector | USB |