Description
|
MPA100 Specifications |
||
|
Operation |
||
| Temperature display | Melting point and melting point range | |
| Start temp. range | (Ambient + 10 °C) to 396 °C, 0 °C min. set point | |
| Stop temp. range | (Start temperature + 4 °C) to 400 °C | |
| Temperature resolution | 0.1 °C | |
| Ramp rate | 0.1 °C to 20 °C per minute (0.1 °C increments) | |
| Heat-up time | 10 minutes (50 °C to 350 °C) | |
| Cool-down time | 10 minutes (350 °C to 50 °C) | |
| Temperature accuracy | ±0.3 °C (up to 100 °C) ±0.5 °C (up to 250 °C) ±0.8 °C (up to 400 °C) |
|
| Reproducibility | 0.2 °C | |
| Temperature sensor | Pt RTD (built-in) | |
| Oven control | Closed-loop PID | |
|
General |
||
| Display | Back-lit, touchscreen LCD (5.8″) | |
| Printer interface | RS-232 serial port. Supports Epson compatible dot-matrix printers. | |
| Computer interface | USB. All instrument functions can be queried and controlled through a high-level command set. | |
| Capillaries | ||
| Dimensions | 1.4 mm to 2.0 mm outside dia., 100 mm length | |
| Capacity | Up to 3 tubes simultaneously | |
| Fill height | 2 mm to 3 mm | |
| Power | 90 to 264 VAC, 47 to 63 Hz, 125 W | |
| Operating temperature | 0 °C to 40 °C, non-condensing | |
| Weight | 9 lbs. | |
| Dimensions | 7.5″ × 10″ × 8.5″ (WHL) | |
| Warranty | One year parts and labor on defects in materials and workmanship | |