The SR555 is a low-noise, high-bandwidth, fixed-gain current (trans-impedance) amplifier designed to work with SRS lockin amplifiers. Current amplifiers provide gain close to the experimental detector, allowing the user to minimize input cable length and its corresponding input capacitance. The SR555 minimizes noise and pickup before they permanently degrade the signal-to-noise ratio, reducing measurement time in noise-limited experiments.

Category:

Description

SR555 Specifications

General
Gain 107 V/A
Bandwidth 120 kHz (-3 dB)
Input noise (typ.) 42 fA/√Hz at 1 kHz
Current input
     Impedance 50 Ω
     Bias current <3 pA
DC bias input
     Range ±5 VDC
     Settling time <150 ms
     Impedance 1 MΩ
Gain accuracy 1 %
Gain stability ±50 ppm/°C
Output 20 Vpp max. balanced differential
10 mA max., 50 Ω
Power Supplied by SR510, SR530, SR810, SR830 or SR850 or SR124 via connector cable
Dimensions 3.0″ × 1.3″ × 5.1″ (WHL)
Weight 10 oz.