Description
Application
The Automated Spectrometer is intended for the detection and analysis of radio nuclides from various types of environmental objects such as rocks, minerals, sludge, slag, soil, plant, sediment and particulate matter in air and water. The spectrometric system is able to determine the composition of a sample based on the photon energy and the activity based on the photon flux. The low-background lead shielding together with the highly pure germanium (HPGe) p-type detector gives precise results even for low activity materials. The fully automated sample changer enables the user to measure up to 40 samples, without having to interact with the Automated Spectrometer. This reliable robotic sample changer increases the productivity and reduces the possibility of health risks for the operator.
COMPLETE SET
- Gamma-ray detector based on HPGe detector in U-style cryostat with remote preamplifier (PA) and 30L Dewar vessel
- Lead shield
- Multichannel analyzer
- Software for spectra processing, identification of radionuclides and calculation of their activities SpectraLineGP
- Spectra analysis software Nuclide Master Plus
- Software for efficiency calibration of arbitrary shaped objects EffMaker
- Automatic sample changer based on robot arm with compact controller
- Table with sample holders and safe cabinet
- Barcode reader together with barcode printer
- Control software for robotic arm
- Master controller with process managing software
- Liquid nitrogen sensor and monitor
- Cable set of 3m length
SPECIFICATION
| Parameter | Value |
|---|---|
| Detection limit for Cs-137 radionuclide specific activity, measurement time 1 hour, Bq/kg | 0.5 |
| Absolute sensitivity to gamma flux for 30%* efficient detector, pulse/quantum | 4.5 x 10-3 |
| Instrumental background intensity for energy range from 40 KeV to 3 MeV, pulse/KeV x sec | 5 x 10-4 |
| Cs-137 radionuclide specific activity measurement error for measurement time 1 hour, % | 20 |
| Shield thickness: Lead wall, mm Copper wall, mm |
100 10 |
| AC power supply: Voltage, V Frequency, Hz |
230 50 |
| * 10% – 160% efficient HPGe detectors are available by separate order. | |
| Lead shield | |
| Thickness of lead walls, mm | 100 |
| Internal dimensions, mm | Ø 202 x 332 |
| Liners | Tin, thickness 1 mm Copper, thickness 9 mm |
| Outer jacket | 1.5 mm low carbon steel |