Description
COMPLETE SET
Basic:
- Liquid scintillation spectrometer TRIEL based on two PMT’s
- Scintillation cocktail (10L)
- Plastic vials 20ml
- Basic software for measurements ASW3L
- User Manual
Accessories:
- Spectrometric software for liquid scintillation spectrometers SpectraDec
- Radionuclide calibration library for H-3, С-14, Cs-137, Sr-90+Y-90
SPECIFICATION
| Parameter | Value |
|---|---|
| Energy range of registered alpha radiation, keV | from 2000 to 10000 |
| Energy range of registered beta radiation, keV | from 0 to 3500 |
| Measurement range of activity of alpha and beta emitting radionuclides, Bq. Expanded uncertainty for measurement time 3600s (k=2) ±10% | from 0,03 to 5∙104 |
| Relative energy resolution for 624 keV of Cs-137, % | < 15 |
| Sensitivity for beta radiation, cps/Bq: – for radionuclide Н-3 – for radionuclide C-14 – for radionuclide Sr-90+Y-90 |
< 0,4 < 0,95 < 0,98 |
| The maximum throughput, cps | > 1∙105 |
| Background intensity in energy range of, cps – H-3 – С-14 – Sr-90+Y-90 |
< 0,3 < 0,5 < 0,7 |
| The minimum measurable activity in sample for measurement time of 30000 s, Bq – H-3 – С-14 – Sr-90+Y-90 |
< 0,02 < 0,013 < 0,007 |
| The instability measurements activity within 24 hours of continuous operation is no more than, % | ±2 |
| Operating temperature range | +10°С … +40°С |
| Relative humidity, % | <80 |
| Power consumption | <40W |
| Power supply | 110 – 240V, 50-60Hz |
| Dimensions, mm | 218x223x473 |
| Weight, kg | 40 |